nanoZ User Manual
14
channels. Increase the pause setting if the ʻtest signal clippedʼ warning is
reported or some channels are not measured (i.e. skipped measurements are
blank and highlighted in yellow in the Report window).
(D) Click the Test probe button to begin the test.
Clicking on Test probe again or pressing ESC at any time will pause the test,
and you will be given the option of continuing the test or stopping.
Impedance test results in the Probe report can be saved by clicking the icon or
selecting File | Save report from the main menu.
!
?)!&%&.+6'9%#+&!/'=&!
This mode cycles through all (or a subset) of the electrodes on the selected
probe, applying a controlled DC constant current to each site. It has two sub-
modes of operation: fixed plating time per channel; and match impedances
mode, whereby the nanoZ will only advance to the next channel when the
electrode site impedance is lowered to the specified Target impedance.
The DC Electroplate mode has many possible applications, including:
o plating with gold or other metals to lower electrode site impedances
o deposition of conductive polymers such as PEDOT to simultaneously
lower the electrode site impedance and improve the site charge capacity
o electrode site cleaning
o in vivo or in vitro rejuvenation of electrode sites
o tissue lesioning and electrode track marking
(A) Select the plating mode, either Fixed plating time, or Match impedances.